Biometer Dissolution Test Automatic Sampling Reciprocating Cylinder Dissolution System
Overview Product Description The RT3-AT automatic sampling reciprocating cylinder dissolution system, which can effectiv
Send your inquiryDESCRIPTION
OverviewBasic Info.
Model NO. | RCY-SZRT-RT3-AT |
Weight | 0-100Kg |
Power Source | AC220V |
Temperature Difference Between Vessels | <±0.5°c |
Adjustable Range of Reciprocating Stroke | 20-100mm |
Thermometric Error | ≤±0.2°c |
Thermometric Resolution | 0.01°c |
Error of Reciprocating Rate | <±5% |
Maximum Sampling Volume | 1-10ml |
Adjustable Range of Reciprocating Rate | 4-60dpm |
Error of Reciprocating Stroke | <±1mm |
MOQ | 1 |
Dissolution Apparatus Level | ≤0.3° |
OEM ODM | Accepted |
Dissolution Vessel Row Number | 7 Tubes*6 Rows / 3 Tubes*3 Rows |
Maximum Running Time | 600 Hours |
Temperature Control Range | Room Temperature - 50°c |
Transport Package | Wooden Box |
Trademark | BIOMETER |
Origin | Shandong |
HS Code | 8479899990 |
Production Capacity | 500 Units/Month |
Product Description
Product DescriptionThe RT3-AT automatic sampling reciprocating cylinder dissolution system, which can effectively provide accurate and precise sampling at each time point, reflecting the advantages of high efficiency, accuracy, stability, easy-to-use software.Features1.Sampling pipe and return pipe are embellish washed before sampling to ensure more accurate sample concentration. 2.After sampling, the residual liquid in the sampling, return and injection needle pipe will be purged and emptied, so that the sampling volume is more accurate and there is no sample residue in the pipe. 3.The functions of automatically repairing the solvent and cleaning the pipeline.
The Teflon sampling pipe with high chemical stability and high anti-adsorption was used to avoid corrosion and sample adsorption.
RT3 provides flexible sampling scheme.Even if there is no automatic sampling workstation, the user can perform the manually sampling through the sampling hole on the dissolution apparatus.
Technical Parameters
Dissolution Apparatus Level | ≤0.3° |
Shaft Verticality | ≤90.0°±0.5° |
Maximum Running Time | 600 Hours |
Dissolution Vessel Row Number | 300ml Standard Vessel: 7 tubes*6 rows1000ml Vessel: 3 tubes*3 rows |
Temperature Control Range | Room Temperature - 50°C |
Thermometric Resolution | 0.01°C |
Thermometric Error | ≤±0.2°C |
Temperature Difference between Vessels | <±0.5°C |
Adjustable Range of Reciprocating Stroke | 20-100mm |
Error of Reciprocating Stroke | <±1mm |
Adjustable Range of Reciprocating Rate | 4-60DPM |
Error of Reciprocating Rate | <±5% |
Sampling Pump | High Precision Compact Injection Pump |
Maximum Settable Sampling Number | 30 Times |
Maximum Sampling Volume | 1-10ml |
Sampling Precision | ≤±1% |
Sample Collector | Sample Collection Quantity: 7 Samples*15 Rows |
Related Products
-
Ase (026) Easy to Operate PMMA/PC Flakes Plastic Recycling with Vibration Drying System
-
Factory Sale Qvc Pneumatic Vacuum Loading Machine
-
Pharmaceutical Gelatin Capsule Softgel Encapsulation Paintball Maker Highly Efficient Soft Gel Gelatin Capsule Making Filling Machine
-
Joston 1000 Liter Agitator Mixing Tank System with Bottom Homogenizer